About this directory
Published studies of wearable light logger performance.
Literature checked 10 November 2026 · 28 studies · 29 device entries · 26 peer reviewed · 2 preprints · 0 conference paper
Using the directory
Browse devices by manufacturer, or sort by most validations, fewest validations or device name. Search for a device, author, title or DOI. The measurement and device-type filters narrow the results.
Each square represents one validation study. Orange hatching identifies developer involvement. Open a device to see its papers, study summaries and source links.
Device outputs shows confirmed photopic illuminance and melanopic EDI features. Examined in studies shows the measurements tested in the matching papers: photopic illuminance, melanopic EDI and angle dependence.
The collection
The directory brings together research publications and preprints that directly test wearable light loggers. Devices are grouped by manufacturer or research developer, with commercial models and research devices identified separately. LYS and LYS Button share one entry.
Papers were found through Europe PMC, searches for named devices and reference lists. Each entry links to the original publication, its DOI and PubMed where available.
Use Submit a validation to suggest a paper or correction by email.
Source and downloads
The site is built with Quarto from a shared literature inventory. The notebooks regenerate the directory, study inventory and downloadable records.
Study table (CSV) · Citations (BibTeX) · Directory notebook · About notebook · GitHub repository
Study inventory
Gemici et al. · 2026 · Preprint
-
Journal / server
-
arXiv
-
Devices
-
ActTrust2; ActLumus
-
Measurements
-
Illuminance; Melanopic EDI
-
Device-specific coverage
-
ActTrust2: Illuminance | ActLumus: Illuminance, Melanopic EDI
-
Setting
-
Field
-
Evidence type
-
Reference comparison
-
Reference
-
JETI Spectraval 1501 in Tübingen; Konica Minolta CL-500A in Izmir.
-
Protocol
-
Fixed tripod rig sampled 433 indoor/outdoor scenes across daylight, artificial and mixed lighting in Tübingen and Izmir. Bootstrap resampling assessed sampling requirements.
-
Findings
-
High association with reference measurements coexisted with systematic underestimation. Bias depended on the lighting condition and scene characteristics. Performance estimates stabilized at approximately 100 sampled scenes in this dataset.
-
Source passage
-
PDF pp. 6–7: Measurement setup; pp. 11–12: Results; Figure 2: A ActLumus mEDI, B ActLumus illuminance, C ActTrust2 illuminance.
DOI: 10.48550/arXiv.2606.20719
Kore et al. · 2026 · Peer reviewed
-
Journal / server
-
npj Biological Timing and Sleep
-
Devices
-
LYS Button; Blue Iris Speck v1.0; Actiwatch Spectrum Plus
-
Measurements
-
Angular response; Illuminance; Melanopic EDI
-
Device-specific coverage
-
LYS Button: Illuminance, Melanopic EDI, Angular response | Blue Iris Speck v1.0: Illuminance, Melanopic EDI, Angular response | Actiwatch Spectrum Plus: Illuminance, Angular response
-
Setting
-
Laboratory + field
-
Evidence type
-
Characterisation
-
Reference
-
Calibrated Gigahertz-Optik BTS256 spectroradiometer; cosine-law angular benchmark.
-
Protocol
-
Three units per model tested under controlled LEDs (0–570 lx) and partly cloudy daylight (42,000–78,000 lx), with separate angular and temperature tests.
-
Findings
-
Speck showed the strongest overall measurement agreement in this comparison. LYS Button overestimated low-light melanopic EDI. Actiwatch had substantial photometric variability. All models showed difficulties under high daylight levels.
-
Source passage
-
Results: Figures 2–3 (illuminance), 4–5 (Speck spectrum), 6 (melanopic EDI), 7 (angular response); Temperature dependence; Discussion limitations.
DOI: 10.1038/s44323-026-00088-y
PubMed record ↗
Pilehchi-Ha et al. · 2026 · Preprint
-
Journal / server
-
medRxiv
-
Devices
-
ActLumus; lido; OcuWEAR
-
Measurements
-
Illuminance; Melanopic EDI
-
Device-specific coverage
-
ActLumus: Illuminance, Melanopic EDI | lido: Illuminance, Melanopic EDI | OcuWEAR: Illuminance, Melanopic EDI
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Gigahertz-Optik CSS-45 reference spectroradiometer.
-
Protocol
-
Comparison under fluorescent, incandescent and multiple LED sources, plus dynamic daylight exposure.
-
Findings
-
Performance depended on both the source spectrum and output metric. Illuminance varied less than melanopic EDI and colour temperature. Bias under dim conditions could change exposure-threshold classification.
-
Source passage
-
Methods and device-specific illuminance, melanopic EDI and CCT comparisons; CCT reported only for lido and OcuWEAR.
DOI: 10.64898/2026.10.06.26364836
Sah et al. · 2026 · Peer reviewed
-
Journal / server
-
npj Biological Timing and Sleep
-
Devices
-
VEET; Clouclip (version not reported); Actiwatch (model not reported); ActLumus; GENEActiv
-
Measurements
-
Illuminance
-
Device-specific coverage
-
VEET: Illuminance | Clouclip (version not reported): Illuminance | Actiwatch (model not reported): Illuminance | ActLumus: Illuminance | GENEActiv: Illuminance
-
Setting
-
Field
-
Evidence type
-
Wearable-to-wearable comparison
-
Reference
-
Concurrent wearables and participant activity logs; no absolute illuminance reference.
-
Protocol
-
Thirty adults wore five sensors for one day; valid light data from 25 participants were compared across illumination categories and with activity logs.
-
Findings
-
Outdoor-photopic categories agreed well between devices, while low-light categories and activity-log estimates differed.
-
Source passage
-
Abstract; Results: Table 2; Introduction (scope); Methods: wearable sensors.
DOI: 10.1038/s44323-026-00113-0
PubMed record ↗
Ishihara et al. · 2025 · Peer reviewed
-
Journal / server
-
SLEEP
-
Devices
-
ActLumus; GENEActiv; ActiGraph wGT3X-BT; LYS Button
-
Measurements
-
Illuminance
-
Device-specific coverage
-
ActLumus: Illuminance | GENEActiv: Illuminance | ActiGraph wGT3X-BT: Illuminance | LYS Button: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
NIST-calibrated ILT560-A spectrometer; free-living comparison between ActLumus and ActiGraph.
-
Protocol
-
Eight units per manufacturer tested under warm/cool LEDs (2–1,000 lx) and daylight (5,000–100,000 lx). Twenty-one adults wore ActLumus and ActiGraph concurrently for one week.
-
Findings
-
ActLumus had the widest tested response range and lowest between-unit variability. LYS showed larger melanopic errors under warm light. GENEActiv and ActiGraph underestimated illuminance; limited sensitivity and saturation produced large differences in daily exposure estimates.
-
Source passage
-
Methods: Laboratory assessment; Results: Figures 2–6 and Free-living comparison (Figure 7).
DOI: 10.1093/sleep/zsaf358
PubMed record ↗
Sarkar et al. · 2025 · Peer reviewed
-
Journal / server
-
Ophthalmic and Physiological Optics
-
Devices
-
Vivior Mark 3+; Clouclip (second generation)
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Vivior Mark 3+: Illuminance | Clouclip (second generation): Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Calibrated Konica Minolta CL-200A chroma meter for illumination; ruler for the separate distance assessment.
-
Protocol
-
Five units of each model attached to spectacles on a mannequin. Light measurements were repeated across seven indoor/outdoor locations on three days. A separate study tested viewing distance in 13 young adults.
-
Findings
-
Both devices tracked the luxmeter; individual differences became more variable above roughly 3,500–4,000 lx. Some bright indoor locations exceeded the 1,000-lx outdoor threshold.
-
Source passage
-
Methods: Study 1 and device descriptions; Results: Validation of ambient illumination measurements, Figures 6–7; Discussion (repeatability).
DOI: 10.1111/opo.13457
PubMed record ↗
Stevenson, Suppiah & Driller · 2025 · Peer reviewed
-
Journal / server
-
Lighting Research & Technology
-
Devices
-
LYS Button
-
Measurements
-
Illuminance
-
Device-specific coverage
-
LYS Button: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Factory-calibrated RS PRO RS-92 digital light meter.
-
Protocol
-
Two LYS Buttons were placed side by side with the reference for eight-hour indoor and outdoor recordings. Separate two-hour recordings assessed between-unit reliability.
-
Findings
-
Buttons read about 25 lx lower indoors and 82 lx higher outdoors on average. Between-unit ICCs were 0.90 indoors and 0.98 outdoors.
-
Source passage
-
Sections 2.1–2.3; Tables 1–2; Section 3.
DOI: 10.1177/14771535251338582
McDowell et al. · 2024 · Peer reviewed
-
Journal / server
-
BMC Biology
-
Devices
-
SpectraWear
-
Measurements
-
Melanopic EDI
-
Device-specific coverage
-
SpectraWear: Melanopic EDI
-
Setting
-
Laboratory
-
Evidence type
-
Calibration study
-
Reference
-
Calibrated SpectroCal spectroradiometer; species-specific photoreceptor weighting functions.
-
Protocol
-
Ten-channel readings were fitted to reference measurements from 13 spectra, then tested at a wider range of irradiances. Performance was assessed for 12 domestic mammal species and humans.
-
Findings
-
Melanopic and rhodopic estimates performed better than short-wavelength cone estimates; limited UV sensitivity constrained some species-specific quantities.
-
Source passage
-
Figure 2; Methods: Light sensor calibration and validation.
DOI: 10.1186/s12915-024-02038-1
PubMed record ↗
Mohammadian et al. · 2024 · Peer reviewed
-
Journal / server
-
IEEE Internet of Things Journal
-
Devices
-
SpectraWear
-
Measurements
-
Angular response; Melanopic EDI
-
Device-specific coverage
-
SpectraWear: Melanopic EDI, Angular response
-
Setting
-
Laboratory + field
-
Evidence type
-
Characterisation + placement
-
Reference
-
SpectroCAL MKII spectroradiometer and calibrated custom spectral source.
-
Protocol
-
Thirteen ten-channel units were calibrated and tested across spectral/intensity conditions, with additional angular-response and body-location experiments.
-
Findings
-
Mean absolute errors in α-opic EDI were below 0.07 log lx; between-unit correlations were at least 0.99 under the reported conditions.
-
Source passage
-
Sections II-D, III-B and IV: calibration, EDI accuracy, directional response and wearing position.
DOI: 10.1109/JIOT.2024.3355330
PubMed record ↗
Stampfli et al. · 2023 · Peer reviewed
-
Journal / server
-
Lighting Research & Technology
-
Devices
-
lido
-
Measurements
-
Angular response; Illuminance
-
Device-specific coverage
-
lido: Illuminance, Angular response
-
Setting
-
Laboratory + field
-
Evidence type
-
Characterisation
-
Reference
-
JETI specbos 1201 spectroradiometer; Ocean Insight STS-VIS for a real-world demonstration.
-
Protocol
-
Integrating-sphere calibration; tests under incandescent, LED, fluorescent and daylight sources; illuminance/CCT and angular-response characterization; near-corneal melanopic irradiance compared with a portable spectroradiometer.
-
Findings
-
The development study reports satisfactory agreement with commercial reference equipment and demonstrates coordinated near-corneal light recording during a field experiment.
-
Source passage
-
Sections 2.5 and 3.1; Figures 8–9 (illuminance/CCT/angular tests); Section 3.3 and Figures 10–11 (melanopic irradiance comparison).
DOI: 10.1177/14771535221147140
PubMed record ↗
Dhakal et al. · 2022 · Peer reviewed
-
Journal / server
-
Ophthalmic and Physiological Optics
-
Devices
-
MyLyt
-
Measurements
-
Illuminance
-
Device-specific coverage
-
MyLyt: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Factory-calibrated Sinometer LX1330B lux meter; activity diaries for indoor/outdoor classification.
-
Protocol
-
Four trackers were tested in a controlled chamber and indoor/outdoor settings. A field phase included 21 adults and eight children.
-
Findings
-
Repeated and between-unit measurements were consistent. Reference agreement varied by setting; diary-based indoor/outdoor classification had high sensitivity and specificity in this sample.
-
Source passage
-
Methods: Phase 1/2 validation; Results: Figures 3–5 and Tables 2–4.
DOI: 10.1111/opo.13061
PubMed record ↗
Bhandari, Mirhajianmoghadam & Ostrin · 2021 · Peer reviewed
-
Journal / server
-
Sensors
-
Devices
-
Clouclip (version not reported); Actiwatch Spectrum Plus
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Clouclip (version not reported): Illuminance | Actiwatch Spectrum Plus: Illuminance
-
Setting
-
Field
-
Evidence type
-
Wearable-to-wearable comparison
-
Reference
-
Concurrent spectacle-mounted Clouclip, wrist Actiwatch Spectrum Plus, daily logs and questionnaire.
-
Protocol
-
Twenty-five adults wore both devices for one week. Illuminance and time above the outdoor threshold were compared with each other and self-reports.
-
Findings
-
Wrist Actiwatch illuminance was systematically lower than Clouclip. Questionnaire estimates of outdoor time exceeded both device estimates.
-
Source passage
-
Sections 2.2–2.3 and light-exposure results.
DOI: 10.3390/s21217096
PubMed record ↗
Falck et al. · 2021 · Peer reviewed
-
Journal / server
-
Journal of Alzheimer's Disease Reports
-
Devices
-
MotionWatch 8
-
Measurements
-
Illuminance
-
Device-specific coverage
-
MotionWatch 8: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Extech LT40 LED light meter.
-
Protocol
-
Seventeen everyday scenarios were sampled for five minutes each, including tests with and without sleeves. Thresholds classified light/dark, electrical/daylight and bright-light therapy.
-
Findings
-
The watch distinguished some lighting classes but had poor absolute agreement and did not reliably identify bright-light therapy. Sleeves commonly reduced readings.
-
Source passage
-
Methods; Table 3; Figures 1–4.
DOI: 10.3233/ADR-200242
PubMed record ↗
Howell et al. · 2021 · Peer reviewed
-
Journal / server
-
Ophthalmic and Physiological Optics
-
Devices
-
Actiwatch 2; Clouclip (second generation)
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Actiwatch 2: Illuminance | Clouclip (second generation): Illuminance
-
Setting
-
Field
-
Evidence type
-
Reference comparison
-
Reference
-
Manufacturer-calibrated Hagner S2 universal photometer.
-
Protocol
-
Devices on a skeleton were compared with a photometer at matched sensor positions in nine indoor/outdoor settings. Five Clouclip M2 units were also compared concurrently.
-
Findings
-
Both devices underestimated illuminance despite high correlations. Clouclip more closely reproduced the reference in the tested range; its between-unit illuminance ICC was 0.85.
-
Source passage
-
Methods: Devices employed and Validity; Results: Tables 2–4, Figures 3–5.
DOI: 10.1111/opo.12860
PubMed record ↗
Mohamed et al. · 2021 · Peer reviewed
-
Journal / server
-
Optics Express
-
Devices
-
Mohamed et al. spectrophotometer
-
Measurements
-
Angular response; Illuminance; Melanopic EDI
-
Device-specific coverage
-
Mohamed et al. spectrophotometer: Illuminance, Melanopic EDI, Angular response
-
Setting
-
Laboratory
-
Evidence type
-
Characterisation
-
Reference
-
Konica Minolta CL-500A calibrated illuminance spectrophotometer.
-
Protocol
-
A pendant/clip spectrophotometer covering 380–780 nm was developed and tested with common light sources; corrections addressed directionality, dark noise, saturation, nonlinearity, stray light and spectral response.
-
Findings
-
After correction, the reported α-opic EDI errors were below 7% for the four tested source types. Optical modification improved angular response; remaining spectral errors differed by source.
-
Source passage
-
Sections 2.2 and 3; Table 1 (angular response), Table 2 and Figure 8 (source comparisons).
DOI: 10.1364/OE.431373
PubMed record ↗
Full text ↗
Nagra et al. · 2021 · Peer reviewed
-
Journal / server
-
Journal of Medical Systems
-
Devices
-
Actiwatch Spectrum Pro
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Actiwatch Spectrum Pro: Illuminance
-
Setting
-
Laboratory
-
Evidence type
-
Repeatability study
-
Reference
-
Controlled Gretag Macbeth Judge II lightbox; comparison among four watches and two sessions.
-
Protocol
-
RGB and white outputs were recorded under simulated daylight D65, illuminant A and cool-white fluorescent light.
-
Findings
-
Significant between-watch differences occurred under all three illuminants; repeated measurements also differed for most watch/channel combinations.
-
Source passage
-
Abstract; Methods: illumination conditions and procedure; Results.
DOI: 10.1007/s10916-020-01694-4
PubMed record ↗
Webler, Chinazzo & Andersen · 2021 · Peer reviewed
-
Journal / server
-
Journal of Physics: Conference Series
-
Devices
-
Spectrace
-
Measurements
-
Angular response; Illuminance
-
Device-specific coverage
-
Spectrace: Illuminance, Angular response
-
Setting
-
Laboratory
-
Evidence type
-
Characterisation
-
Reference
-
Reference spectra; LMT Pocket-Lux 2 for angular-response measurements.
-
Protocol
-
Prototype tests examined spectral reconstruction, angular response and intensity linearity; a separate survey assessed wearability.
-
Findings
-
Measured angular mismatch was 2.07%. Spectral reconstruction and intensity response depended on the tested signal and source conditions.
-
Source passage
-
Sections 2.1 and 3.1.1–3.1.3; Figures 1–4.
DOI: 10.1088/1742-6596/2042/1/012120
Wen et al. · 2021 · Peer reviewed
-
Journal / server
-
Acta Ophthalmologica
-
Devices
-
Clouclip (first generation)
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Clouclip (first generation): Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Konica Minolta T-10 illuminance meter.
-
Protocol
-
Fifteen first-generation units were tested in 28 light environments spanning 50–38,000 lx, with repeated measurements. A separate participant study compared device records with questionnaires.
-
Findings
-
Controlled light tests reported a maximum relative error of 2.23% and high within- and between-unit consistency.
-
Source passage
-
Methods: Validation of Clouclip; light-intensity validation results.
DOI: 10.1111/aos.14785
PubMed record ↗
Bhandari & Ostrin · 2020 · Peer reviewed
-
Journal / server
-
Ophthalmic and Physiological Optics
-
Devices
-
Clouclip (second generation)
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Clouclip (second generation): Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Dr. Meter LX1330B digital lux meter.
-
Protocol
-
Clouclip and reference sensors were aligned in indoor and outdoor locations, including shade and direct sunlight. The paper separately tested viewing distance.
-
Findings
-
Illuminance was strongly correlated with the reference, but disagreement grew at high levels, particularly above 20,000 lx.
-
Source passage
-
Methods: Ambient illumination; Results: Figure 7.
DOI: 10.1111/opo.12735
PubMed record ↗
Joyce, Zele, Feigl & Adhikari · 2020 · Peer reviewed
-
Journal / server
-
Journal of Sleep Research
-
Devices
-
GENEActiv Original; Actiwatch 2
-
Measurements
-
Illuminance
-
Device-specific coverage
-
GENEActiv Original: Illuminance | Actiwatch 2: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Reference comparison
-
Reference
-
Calibrated laboratory-standard photometer.
-
Protocol
-
Compared device illuminance with a photometer under white LED light and natural sunlight; also tested neutral-density filters to extend measurement range.
-
Findings
-
At about 20,000 lux LED light, GENEActiv and Actiwatch 2 recorded about 50% and 25% of reference illuminance, respectively. At about 30,000 lux sunlight, GENEActiv saturated and Actiwatch 2 recorded about 46%.
-
Source passage
-
Abstract: photometer comparison, LED/sunlight under-reading, saturation and neutral-density filters.
DOI: 10.1111/jsr.12963
PubMed record ↗
Stone, McGlashan, Facer-Childs, Cain & Phillips · 2020 · Peer reviewed
-
Journal / server
-
Clocks & Sleep
-
Devices
-
GENEActiv Original
-
Measurements
-
Illuminance
-
Device-specific coverage
-
GENEActiv Original: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Characterisation
-
Reference
-
J17 Luma Color illuminance meter; MK350N spectrometer for spectral measurements.
-
Protocol
-
Ten units tested under fluorescent and LED illumination, darkness, overcast outdoor daylight, narrowband blue/green/red LEDs and an infrared camera illuminator.
-
Findings
-
Outputs tracked illuminance but generally underestimated it, particularly below 100 lux. Accuracy depended on spectrum: green light was under-reported, red light over-reported, and infrared affected the measurements.
-
Source passage
-
§2.1–2.4: devices, stimuli and protocol; Table 2 and Results: intensity accuracy, spectral/infrared effects and inter-device reliability.
DOI: 10.3390/clockssleep2020012
PubMed record ↗
Argüelles-Prieto et al. · 2019 · Peer reviewed
-
Journal / server
-
Frontiers in Physiology
-
Devices
-
Kronowise
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Kronowise: Illuminance
-
Setting
-
Laboratory
-
Evidence type
-
Calibration study
-
Reference
-
Ocean Optics spectroradiometer with CC-3 cosine corrector; CIE S 026 calculations from reference spectra.
-
Protocol
-
Commercial lamps and sunlight were compared with three-channel Kronowise readings; separate calibration relationships were developed for natural and two artificial-light classes.
-
Findings
-
Source-specific transformations improved correspondence with reference visible and circadian-light estimates relative to raw outputs.
-
Source passage
-
Methods: calibration; Results: light-source classification and calibrated visible/circadian measurements.
DOI: 10.3389/fphys.2019.00822
PubMed record ↗
Ostrin · 2017 · Peer reviewed
-
Journal / server
-
Optometry and Vision Science
-
Devices
-
Actiwatch Spectrum
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Actiwatch Spectrum: Illuminance
-
Setting
-
Laboratory + field
-
Evidence type
-
Calibration study
-
Reference
-
Lux meter and UV sensor; models unresolved in the inspected abstract.
-
Protocol
-
The paper reports indoor/outdoor device calibration in addition to 14-day recordings in 55 adults and comparisons with activity questionnaires.
-
Findings
-
An original calibration component is reported alongside the exposure study; numerical sensor agreement has not been extracted from the accessible abstract.
-
Source passage
-
Publisher abstract, Methods: calibration against a lux meter and UV sensor.
DOI: 10.1097/OPX.0000000000001013
PubMed record ↗
Price, Lyachev & Khazova · 2017 · Peer reviewed
-
Journal / server
-
Journal of the Optical Society of America A
-
Devices
-
MotionWatch 8; ActTrust; GENEActiv Original; Actiwatch 2
-
Measurements
-
Angular response
-
Device-specific coverage
-
MotionWatch 8: Angular response | ActTrust: Angular response | GENEActiv Original: Angular response | Actiwatch 2: Angular response
-
Setting
-
Laboratory
-
Evidence type
-
Characterisation
-
Reference
-
Photopic and photoreceptor sensitivity functions, cosine response and reference values for dynamic characterization.
-
Protocol
-
New optical tests and existing evidence were combined. This record indexes four complete-protocol models with individually identifiable spectral and angular results: MotionWatch 8, ActTrust, GENEActiv Original and Actiwatch 2.
-
Findings
-
Optical characterization covers directional and spectral sensitivity, linearity, dynamic range and resolution, with a proposed assessment protocol. Results are model- and channel-specific.
-
Source passage
-
Table 1: complete-protocol models; Table 2: spectral mismatch; Table 3: cosine-response error; Table 4: dynamic performance.
DOI: 10.1364/JOSAA.34.000545
PubMed record ↗
Markvart et al. · 2015 · Peer reviewed
-
Journal / server
-
LEUKOS
-
Devices
-
Actiwatch Spectrum
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Actiwatch Spectrum: Illuminance
-
Setting
-
Laboratory
-
Evidence type
-
Between-unit calibration
-
Reference
-
Side-by-side comparison among 48 Actiwatch Spectrum units.
-
Protocol
-
RGB and white outputs were compared under different light sources to develop a field-calibration procedure.
-
Findings
-
The study documents between-unit response variation and proposes corrections based on concurrent measurements.
-
Source passage
-
Author-deposited abstract: study aim, device model and protocol.
DOI: 10.1080/15502724.2015.1020948
Figueiro, Hamner, Bierman & Rea · 2013 · Peer reviewed
-
Journal / server
-
Lighting Research & Technology
-
Devices
-
Daysimeter-S; Daysimeter-D; Actiwatch Spectrum
-
Measurements
-
Angular response; Illuminance
-
Device-specific coverage
-
Daysimeter-S: Illuminance, Angular response | Daysimeter-D: Illuminance, Angular response | Actiwatch Spectrum: Illuminance, Angular response
-
Setting
-
Laboratory + field
-
Evidence type
-
Characterisation
-
Reference
-
Photopic V(λ), cosine response and calibrated near-eye Daysimeter measurements for placement comparisons.
-
Protocol
-
Optical characterization and field comparisons: 12 adults over 65 wore Daysimeters at four body locations and a wrist Actiwatch; five days analyzed.
-
Findings
-
Daysimeters performed better photometrically than Actiwatch Spectrum. Wrist measurements differed from near-cornea exposure.
-
Source passage
-
§2.2.1, Figures 3–5: optical performance; §3.2–3.3, Figure 6: Daysimeter placement; Figure 7: wrist comparison.
DOI: 10.1177/1477153512450453
PubMed record ↗
Price, Khazova & O’Hagan · 2012 · Peer reviewed
-
Journal / server
-
Lighting Research & Technology
-
Devices
-
Actiwatch Spectrum
-
Measurements
-
Angular response
-
Device-specific coverage
-
Actiwatch Spectrum: Angular response
-
Setting
-
Laboratory
-
Evidence type
-
Characterisation
-
Reference
-
Optical characterization; abstract does not identify all reference instruments.
-
Protocol
-
Sixteen Actiwatch Spectrum units were evaluated for spectral response, directional response and dynamic range.
-
Findings
-
The paper reports optical performance and proposes calibration, deployment and analysis techniques for circadian studies.
-
Source passage
-
Abstract: 16 units; spectral response, directional response and dynamic range.
DOI: 10.1177/1477153511433171
Bierman, Klein & Rea · 2005 · Peer reviewed
-
Journal / server
-
Measurement Science and Technology
-
Devices
-
Daysimeter (2005 prototype)
-
Measurements
-
Illuminance
-
Device-specific coverage
-
Daysimeter (2005 prototype): Illuminance
-
Setting
-
Laboratory
-
Evidence type
-
Characterisation
-
Reference
-
Photo Research LRS 450 Light Standard; spectral matching against the photopic response.
-
Protocol
-
Two prototypes were checked for spectral response. Linearity was tested against an integrating-sphere source with an approximately illuminant-A spectrum.
-
Findings
-
Photopic spectral-mismatch values were 3.8% and 5.4%. The tested response departed from linearity by a few percent across roughly 3.5 decades of irradiance.
-
Source passage
-
Author PDF, Section 3.1.3: Radiation measurement verification; Figure 3.
DOI: 10.1088/0957-0233/16/11/023